Are semi-automated software program designed for adults accurate for the identification of vertebral fractures in children?

Eur Radiol. 2019 Dec;29(12):6780-6789. doi: 10.1007/s00330-019-06250-4. Epub 2019 May 22.

Abstract

Objectives: To assess whether diagnostic accuracy of morphometric vertebral fracture (VF) diagnosis in children can be improved using AVERT™ (a 33-point semi-automated program developed for VF diagnosis in adults) compared with SpineAnalyzer™ (a 6-point program), which has previously been shown to be of insufficient accuracy.

Materials and methods: Lateral spine radiographs (XR) and dual-energy X-ray absorptiometry (DXA) scans of 50 children and young people were analysed by two observers using two different programs (AVERT™ and SpineAnalyzer™). Diagnostic accuracy (sensitivity, specificity, false-negative (FN) and false-positive rates (FP)) was calculated by comparing with a previously established consensus arrived at by three experienced paediatric musculoskeletal radiologists, using a simplified algorithm-based qualitative scoring system. Observer agreement was calculated using Cohen's kappa.

Results: For XR, overall sensitivity, specificity, FP and FN rates using AVERT™ were 36%, 95%, 5% and 64% respectively and 26%, 98%, 2% and 75% respectively, using SpineAnalyzer™. For DXA, overall sensitivity, specificity, FP and FN rates using AVERT™ were 41%, 91%, 9% and 59% respectively and 31%, 96%, 4% and 69% respectively, using SpineAnalyzer. Reliability (kappa) ranged from 0.34 to 0.37 (95%CI, 0.26-0.46) for AVERT™ and from 0.26 to 0.31 (95%CI, 0.16-0.44) for SpineAnalyzer™. Inter- and intra-observer agreement ranged from 0.41 to 0.47 for AVERT™ and from 0.50 to 0.79 for SpineAnalyzer™.

Conclusion: AVERT™ has slightly higher accuracy but lower observer reliability for the representation of vertebral morphometry in children when compared with SpineAnalyzer™. However, neither software program is satisfactorily reliable for VF diagnosis in children.

Key points: • SpineAnalyzer™ and AVERT™ have low diagnostic accuracy and observer agreement when compared to three paediatric radiologists' readings for the diagnosis of vertebral fractures (VF) in children. • Neither AVERT™ nor SpineAnalyzer™ is satisfactorily reliable for VF diagnosis in children. • Development of specific paediatric software and normative values (incorporating age-related physiological variation in children) is required.

Keywords: Children; DXA scan; Osteoporosis; Paediatric; Vertebral.

MeSH terms

  • Absorptiometry, Photon / methods*
  • Adolescent
  • Algorithms
  • Child
  • Child, Preschool
  • Female
  • Humans
  • Male
  • Observer Variation
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spinal Fractures / diagnostic imaging*
  • Spinal Injuries / diagnostic imaging
  • Spine / diagnostic imaging